Anandamide Aea Elisa Test Kit 0.022ng / ml With Strong Specificity / Sensitivity

China Shanghai Korain Biotech Co., Ltd certification
China Shanghai Korain Biotech Co., Ltd certification
Providing high quality products、perfect after-sales service, and try best to meet the requirement of customers. We are very willing to continue long-term cooperation.

—— Tommy

We have confidence in working with you to expand the market one more year.

—— Lilis

I'm Online Chat Now

Anandamide Aea Elisa Test Kit 0.022ng / ml With Strong Specificity / Sensitivity

Anandamide Aea Elisa Test Kit 0.022ng / ml With Strong Specificity / Sensitivity
Anandamide Aea Elisa Test Kit 0.022ng / ml With Strong Specificity / Sensitivity

Large Image :  Anandamide Aea Elisa Test Kit 0.022ng / ml With Strong Specificity / Sensitivity

Product Details:

Place of Origin: Shanghai, China
Brand Name: BT Lab
Certification: CE, ISO9001:2005, MSDS
Model Number: Cat.No E3875Hu

Payment & Shipping Terms:

Minimum Order Quantity: Negotiation
Price: Negotiation
Packaging Details: Wrapped with ice pack and styrofoam package
Delivery Time: 1-3 business days, bulk order within one week
Supply Ability: Western Union, T/T
Detailed Product Description
Custom: Available Discount: Available
Test Method: Sandwich Organism Species: Human
OEM: Acceptable Cat.No: E3875Hu
High Light:

enzyme immunoassay kit

,

enzyme assay kits

Human High Specificity and Sensitivity Anandamide AEA ELISA Kit 96 Wells

 

Cat.No E3875Hu

Standard Curve Range: 0.05ng/ml - 20ng/ml

Sensitivity: 0.022ng/ml

Size: 96 wells

Storage: Store the reagents at 2-8°C. For over 6-month storage refer to the expiration date keep it at -20°C. Avoid repeated thaw cycles. If individual reagents are opened it is recommended that the kit be used within 1 month.

*This product is for research use only, not for use in diagnosis procedures. It’s highly recommend to read this instruction entirely before use.

 

Intended Use

This sandwich kit is for the accurate quantitative detection of Human Anandamide (also known as AEA) in serum, plasma, cell culture supernates, cell lysates, tissue homogenates.

 

Reagent Provided

Components Quantity
Standard Solution (24ng/ml) 0.5ml x1
Pre-coated ELISA Plate 12 * 8 well strips x1
Standard Diluent 3ml x1
Streptavidin-HRP 6ml x1
Stop Solution 6ml x1
Substrate Solution A 6ml x1
Substrate Solution B 6ml x1
Wash Buffer Concentrate (30x) 20ml x1
Biotinylated Human AEA Antibody 1ml x1
User Instruction 1
Plate Sealer 2 pics
Zipper bag 1 pic

 

Material Required But Not Supplied

  • 37°C±0.5°C incubator
  • Absorbent paper
  • Precision pipettes and disposable pipette tips
  • Clean tubes
  • Deionized or distilled water
  • Microplate reader with 450 ± 10nm wavelength filter

 

Precautions

  • Prior to use, the kit and sample should be warmed naturally to room temperature 30 minutes.
  • This instruction must be strictly followed in the experiment.
  • Once the desired number of strips has been removed, immediately reseal the bag to protect the remain from deterioration. Cover all reagents when not in use.
  • Make sure pipetting order and rate of addition from well-to-well when pipetting reagents.
  • Pipette tips and plate sealer in hand should be clean and disposable to avoid cross-contamination.
  • Avoid using the reagents from different batches together.
  • Substrate solution B is sensitive to light, don’t expose substrate solution B to light for a long time.
  • Stop solution contains acid. Please wear eye, hand and skin protection when using this material. Avoid contact of skin or mucous membranes with kit reagent.
  • The kit should not be used beyond the expiration date.

 

Reagent Preparation

All reagents should be brought to room temperature before use.

Standard Reconstitute the 120μl of the standard (24ng/ml) with 120μl of standard diluent to generate a 12ng/ml standard stock solution. Allow the standard to sit for 15 mins with gentle agitation prior to making dilutions. Prepare duplicate standard points by serially diluting the standard stock solution (12ng/ml) 1:2 with standard diluent to produce 6ng/ml, 3ng/ml, 1.5ng/ml and 0.75ng/ml solutions. Standard diluent serves as the zero standard(0 ng/ml). Any remaining solution should be frozen at -20°C and used within one month. Dilution of standard solutions suggested are as follows:

 

12ng/ml Standard No.5 120μl Original Standard + 120μl Standard Diluent
6ng/ml Standard No.4 120μl Standard No.5 + 120μl Standard Diluent
3ng/ml Standard No.3 120μl Standard No.4 + 120μl Standard Diluent
1.5ng/ml Standard No.2 120μl Standard No.3 + 120μl Standard Diluent
0.75ng/ml Standard No.1 120μl Standard No.2 + 120μl Standard Diluent

 

Standard Concentration Standard No.5 Standard No.4 Standard No.3 Standard No.2 Standard No.1
24ng/ml 12ng/ml 6ng/ml 3ng/ml 1.5ng/ml 0.75ng/ml

 

Wash Buffer Dilute 20ml of Wash Buffer Concentrate 30x into deionized or distilled water to yield 500 ml of 1x Wash Buffer. If crystals have formed in the concentrate, mix gently until the crystals have completely dissolved.

 

Assay Procedure

1. Prepare all reagents, standard solutions and samples as instructed. Bring all reagents to room temperature before use. The assay is performed at room temperature.

2. Determine the number of strips required for the assay. Insert the strips in the frames for use. The unused strips should be stored at 2-8°C.

3. Add 50μl standard to standard well. Note: Don’t add antibody to standard well because the standard solution contains biotinylated antibody.

4. Add 40μl sample to sample wells and then add 10μl anti-AEA antibody to sample wells, then add 50μl streptavidin-HRP to sample wells and standard wells ( Not blank control well ). Mix well. Cover the plate with a sealer. Incubate 60 minutes at 37°C.

5. Remove the sealer and wash the plate 5 times with wash buffer. Soak wells with at least 0.35 ml wash buffer for 30 seconds to 1 minute for each wash. For automated washing, aspirate all wells and wash 5 times with wash buffer, overfilling wells with wash buffer. Blot the plate onto paper towels or other absorbent material.

6. Add 50μl substrate solution A to each well and then add 50μl substrate solution B to each well. Incubate plate covered with a new sealer for 10 minutes at 37°C in the dark.

7. Add 50μl Stop Solution to each well, the blue color will change into yellow immediately.

8. Determine the optical density (OD value) of each well immediately using a microplate reader set to 450 nm within 10 minuets after adding the stop solution.